For Displacement sensors, Resistive output sensors, etc. |
Setting range |
Measurement range |
Open circuit voltage |
Measuring current |
Measurement error rate |
Minimum resolution |
150Ω | 0-160Ω | 3.3kV | 1000μA | ±0.02%FS | 0.01Ω |
1.5kΩ | 0-1.6kΩ | 0.1Ω | |||
10kΩ | 0-10kΩ | 100μA | 0.5Ω | ||
40kΩ | 0-40kΩ | 50μA | 1.5Ω |
Measurements
ApplicationsCustomization is possible, built into box, multi-CH, etc.
>Customization page Can be calibrated as an embedded product. Calibration documents and traceability charts can be issued.
|
CAN output [Option]
|
16bit analog output [Option] Alarm output [Option] USB Type-C [standard feature] |
RR-2401 | Display only, Equipped with USB communication | 32,500 yen and up* [tax not included] |
RR-2451 | DIN rail mounting, no display | 28,100 yen and up* [tax not included] |
CSV logging | Compare and match set value lists | PC save and copy of set values | Prevention of misoperation | Quickly set up multiple units | ||||
Time-specified logging, such as 600 seconds, is also available. Custom logger software is also available. | Multiple lists of set values can be compared and collated. | It prevents errors in note-taking and manual entry, and reduces management effort and costs. | Settings and confirmations can be made on the PC screen, preventing misoperation of the main unit's buttons. | Even multiple units can be easily configured by simply writing the configuration data. |
USB Type-C connector standard: configuration, logging, power supply | Setup unit without LINKSHIP or PC required: Program Loader PL-5A [sold separately] PL-5A Product Page *Setting only / data logging and power supply not available |
|
Rear USB 2411 type also available. |
Simply connect the wires to the connector and plug it in by hand. Detachable Plug Connectors |
Fixed Connectors [Customized support]non-detachable |
Mounting hooks: Fixing the main body |
||||
Spring lock type [standard] |
M2 screw type [J04 option] |
|
*Not included in 2431 and 2451 types. |
Carry and measure | Built in panel | DIN rail mounting | |||||||||||
Carry PC, instrument, and cable, and measure with USB power supply. Storage case BXS-1 is also available. | 24x48 size contributes to downsizing | One-touch fixture and display-less type are available. | |||||||||||
|
RR-24 Dimensions |
|
Applications | See more Applications > |
Valve opening measurement and monitoring
Resistor to CAN conversion etc.
Specifications
Name Model |
Resistance meter RR-2401 |
||||||
Input | Number of input points |
1 point | |||||
Setting range 150Ω 1.5kΩ 10kΩ 40kΩ |
Measurement range 0 - 160Ω 0 - 1.6kΩ 0 - 10kΩ 0 - 40kΩ |
Open circuit voltage 3.3kV 3.3kV 3.3kV 3.3kV |
Measuring Current 1000μA 1000μA 100μA 50μA |
Measurement error rate ±0.02%FS ±0.02%FS ±0.02%FS ±0.02%FS |
Minimum Resolution 0.01Ω 0.1Ω 0.5Ω 1.5Ω |
||
Measurement range Sensor Power supply - output [Option] |
150Ω / 1.5kΩ / 10kΩ / 40kΩ [Ambient temperature 23°C] [H] DC +12V±5% 60mA max / [L] DC +5V±5% 150mA max *When USB power is supplied, power for the sensor is not output. |
||||||
Calculation | Measuring method Moving Average Number Hold function Hold input Additional control input Setting value memory |
ΔΣ integral method 1 to 99 Set from current value, max. value, min. value, max. fluctuation range, zero shift (hold / external control input) Zero shift function, set the current input level as the zero reference value for measurement ON when shorted to 1-point GND, OFF when open (hold function is set) ON when shorted to 2-point GND, OFF when open (hold function assigned and set for each input) Nonvolatile memory (EEPROM) |
|||||
Display | Display Color Indicator (Numerical Display) Decimal point position Display update time Display moving average times Indicator |
[Standard] RR-2401 Red LED / [Option] RR-2401-G Green LED 5-digit, 7-segment LED Character height 9mm / Display range:-19999 to 99999 Over indication: OL / Zero indication: Leading zero suppress Semi-fixed Decimal point position is set in program mode. [0] *****, [1] ****.* to [4] *.**** 0.3 sec (settable from 0.1 to 9.9 sec) 1 time (settable from 1 to 9 times) LED 1 point (selected from power/ control input) |
|||||
USB communication |
Communication specification Communication function Output update time |
USB2.0 (Serial port communication by USB virtual COM 230.4kbps) Writing/reading of set values/continuous output of measured values 1 to 9999ms, settable in 1ms increments |
|||||
General Information |
Operating temp./humidity Power supply voltage Power consumption Isolation Withstand voltage Weight |
-10°C to +50°C / 35 to 85%RH (non-condensing) 7 to 60V DC with reverse connection protection. Power supply from USB host: 5V 3W or less Power supply / Resistance meter input / Other input/output Between each terminal of the above isolations: 500V DC between each terminal for 1 minute Approx. 60g |
|||||
CAN Option | Communication specifications Communication function Setting item Output update time |
CAN2.0B 10 k / 20 k / 50 k / 100 k / 250 k / 500 k / 1M bps, [Format] INTEL/MOTOROLA, [Frame] 11bit/29bit Writing/reading of set values/continuous output of measured values CAN transmit ID, CAN receive ID 1 to 9999 ms (settable in 1 ms increments) |
|||||
RS-485 communication option |
Number of connected units Communication format Communication method Communication standard Communication speed Data format Communication code Communication function |
32 units (max.) 2-wire multi-drop serial communication Asynchronous RS-485 9,600 /19,200 /38,400bps Set in program mode Start bit: 1 bit / Stop bit: 1 bit / Data length: 8 bits / Parity bit: None ASCII Writing/reading set values/measurement value output (response to request command) Modbus RTU support is available upon request. |
|||||
Analog output option |
Output signals D/A conversion method Output resolution Load resistance Output accuracy Temperature fluctuation Output scaling Output update time Response time after calculation |
[E1] 0 to 10 V, [E5] 0 to 5 V, [R1] ±10 V, [R5] ±5 V, [I1] 4 to 20 mA [Select a model] DAC conversion method 16bit (50,000 or more) [Voltage output] 4.7kΩ or more / [Current output] 300Ω or less [Voltage output] ±0.1% of FS @ 23°C / [Current output] ±0.1% of FS @ 23°C ±200ppm/℃ or less Arbitrary scaling is possible by full scale and zero scale settings 1 to 9999ms settable in 1ms increments 1 ms or less, 0 - 90% response |
|||||
Alarm [Comparator] output option |
Setting method Number of outputs Output form Output load voltage Output logic Update time |
Set in program mode 2 points Isolated non-contact output (photo-MOS relay) No polarity. When comparator operating condition is ON, resistance between COMP and COM: 50 Ω or less Peak AC/DC280V 100mA (resistive load) / ON resistance 50Ω or less Comparison is made with polarity (0 is greater than minus) [Over] ON when set value or more / [Under] ON when set value or less/ [Within range] ON within the set value, OFF outside the range [Out of range] ON outside the set value, OFF inside the range Depends on analog output update time setting. Response time 1ms (max) |