For Torque sensors, Density sensors, etc. with frequency deviation output |
Customization - built into box, multi-CH, etc. | Read more > |
CAN output 1ms update [Option]
|
16bit analog output 0.1ms update [Option] Alarm output 1ms update [Option] USB Type-C 1ms update [standard feature] |
Input 600μHz - 1MHz, resolution 10.4ns
Supports signals | 5V Logic | zero cross | NPN open collector | 12V Logic | Customized Signals |
Various options | Input Isolation | Line driver input | [Please contact us for more than 1 MHz] | ||
Flexible adjustment | Trigger Level 0-9.99V | Hysteresis Select | Pull-up ON/OFF |
PERIOMATICTM calculation High accuracy from ultra-low speed to high speed. Follows sudden changes and stops. |
Frequency measurement technology originally developed by COCORESEARCH |
Front operation. Setting and measuring
(A) Operation indicator LED (B) Front sheet button - [In program mode] MODE, digit selection, numeric setting, ENTER - [In measurement mode] Span shift, Zero shift - Key lock: Prevents unintentional operation by field workers. |
Blue/pink type [Option] *Designs are subject to change. |
Smart configuration and logging | LINKSHIP Page > |
CSV logging | Compare and match set value lists | PC save and copy of set values | Prevention of misoperation | Quickly set up multiple units | ||||
Time-specified logging, such as 600 seconds, is also available. Custom logger software is also available. | Multiple lists of set values can be compared and collated. | It prevents errors in note-taking and manual entry, and reduces management effort and costs. | Settings and confirmations can be made on the PC screen, preventing misoperation of the main unit's buttons. | Even multiple units can be easily configured by simply writing the configuration data. |
USB Type-C connector standard | Setup unit without LINKSHIP or PC required: Program Loader PL-5A [sold separately] |
PL-5A Product Page *Setting only / data logging and power supply not available |
Applications | See more Applications > |
TQ-4801 |
Torque → |
PLC, Logger, PC, etc. |
Setting Example [A] Torque ±400N・m [B] Central frequency 60000Hz [C] Deviation frequency ±20000Hz Input-Output Relationship
|
|||||||||||
TQ-48401 Speedmeter/ Counter [Switchable] |
Speed → |
TQ-4801 |
CAN USB Analog output Alarm output → |
PLC, Logger, PC, etc. |
Setting Example [A] Mixing ratio ±50% [B] Central frequency 100Hz [C] Deviation frequency ±50Hz Input-Output Relationship
|
Specifications
Name Model Measuring method |
Torque meter [frequency deviation meter] TQ-4801 PeriomaticTMB method Cycle calculation method |
|||
Input | Number of input points Input frequency Input resolution Input Signal Input method |
1 point 0.0006Hz to 1.0MHz Approx. 10.4ns (96MHz) 5V logic, zero cross, NPN open collector, 12V logic, input customized signal Single-phase signal |
||
[*] General-purpose input option |
Input Level and Sensitivity | 5V logic Zero-cross NPN open collector 12V logic Customized input |
Trigger level 2.5V 0V 2.5V 6V 0-9.0V in 10mV units |
Hysteresis ±0.4V ±0.03V ±0.4V ±0.4V Small:±0.03V, Medium : 0.4V, Large: ±1.3V, trigger level 4.7V or lower/ ±2.6V, trigger level 4.7V or higher |
Input pulse Input resistance Input withstand voltage |
[Small] [Medium] 0.2 μs or more [Large] 0.3 μs or more, for both H and L levels 5V logic, zero cross, NPN open collector, and Input customized signal (trigger level 4.7V or lower): 10kΩ 12V logic, input customized signal (trigger level 4.71V or higher): 5kΩ ±30V *Inputs and outputs are not insulated for the general-purpose input option. |
|||
[D] Line driver input option |
Input signal Input withstand voltage Recommended line driver Termination resistance |
Line driver signal ±25V (differential voltage) Equivalent to AM26LS31 300Ω (standard specification) *Input and output are isolated for the line driver input option. |
||
Input low-pass filter Trigger edge Input withstand voltage Input connector Sensor Power supply output[Option] |
Select from OFF, 500Hz, 5kHz, 120kHz, or 800kHz Fall ±30V Terminal block 7.62mm pitch M3 free terminal screw [H] DC +12V±5% 60mA max / [L] DC +5V±5% 150mA max |
|||
Calculation | Measurement mode Calculation rate Frequency divider ratio Output moving average Dynamic PredictiveTM Hold function Additional control input Low cut function Dual range |
Custom (Torque) / Frequency meter / Period meter (seconds only) Center frequency, deviation frequency at rating, rated torque, and output polarity are set for input frequency. 1 to 60000 1 to 99 8 levels (including continuous and stop prediction) 1 point, Set from current value, maximum value, minimum value, maximum fluctuation range (Hold input, Additional control input) ON when shorted to GND, OFF when open (hold function is set) 2 points, ON by short-circuiting to GND, OFF by opening (hold function assigned and set for each input) Low cut function Stop judgment at a speed lower than the set value Two different calculation rates can be set *Selected by hold input according to hold function assignment |
||
Display | Display Color Indicator (Numerical Display) Decimal point position Display update time Display accuracy Indicator (3 round LEDs) |
[Standard] TQ-4801 Red LED / [Option] TQ-4801-G Green LED 6-digit, 7-segment LED Character height 14.2mm / Display range:-199999 to 999999 Polarity indication: Reverse rotation -Lighted / Over indication: OL / Zero indication: Leading zero suppress Semi-fixed Decimal point position is set in program mode. [0] ******, [1] *****.* to [4] **.**** 0.3 sec (settable from 0.1 to 9.9 sec) 5ppm ± 1 digit - IND Pulse Input Indicator [Single-phase pulse] Flashes at falling edge of pulse High-speed pulse lights up continuously [UP/DOWN signal] Flashes at falling edge of UP/DOWN signal. Flashing at falling edge of UP/DOWN signal - HI: HIGH comparator output ON - LO: LOW comparator output ON |
||
USB communication |
Communication specification Communication function Output update time |
USB2.0 (Serial port communication by USB virtual COM 230.4kbps) Writing/reading of set values/continuous output of measured values 1 to 9999.9ms, settable in 0.1ms increments (No output when output update time is set to 1 to 9) |
||
General Information |
Operating temp./humidity Power supply voltage Power consumption Isolation Withstand voltage Weight |
-10°C to +50°C / 35 to 85%RH (non-condensing) TQ-4801-AF: AC100-240V (Maximum rated AC85-264V) 50/60Hz TQ-4801-DF: 12-30V DC with reverse connection protection 3W or less [General-purpose input type]: Power supply / Other input/output [Line driver input type]: Power supply / Frequency signal input / Other input/output Between each terminal of the above isolations: 500V DC between each terminal for 1 minute Approx. 340g |
||
CAN Option | Communication specifications Communication function Setting item Output update time |
CAN2.0B 10 k / 20 k / 50 k / 100 k / 250 k / 500 k / 1M bps, [Format] INTEL/MOTOROLA, [Frame] 11bit/29bit Writing/reading of set values/continuous output of measured values CAN transmit ID, CAN receive ID 1 to 9999.9 ms (settable in 0.1 ms increments) (When output update time is set to 1 to 9, instantaneous values are output every 1 ms.) |
||
RS-485 communication option |
Number of connected units Communication format Communication method Communication standard Communication speed Data format Communication code Communication function |
32 units (max.) 2-wire multi-drop serial communication Asynchronous RS-485 9,600 /19,200 /38,400bps Set in program mode Start bit: 1 bit / Stop bit: 1 bit / Data length: 8 bits / Parity bit: None ASCII Writing/reading set values/measurement value output (response to request command) Modbus RTU support is available upon request. |
||
Analog output option |
Output signals D/A conversion method Output resolution Load resistance Output accuracy Temperature fluctuation Output scaling Output update time Input/output delay time |
[E1] 0 to 10 V, [E5] 0 to 5 V, [R1] ±10 V, [R5] ±5 V, [I1] 4 to 20 mA [Select a model] DAC conversion method 16bit (50,000 or more) [Voltage output] 4.7kΩ or more / [Current output] 300Ω or less [Voltage output] ±0.1% of FS @ 23°C / [Current output] ±0.1% of FS @ 23°C ±200ppm/℃ or less Arbitrary scaling is possible by full scale and zero scale settings 0.1 to 9999.9ms settable in 0.1ms increments Output update time + analog rise time 150 μs max. |
||
Alarm [Comparator] output option |
Setting method Number of outputs Output form Output load voltage Output logic Update time |
Set in program mode 2 points (or 1 point) Depends on model selection conditions. 1 point for line driver input and analog output with comparator output specification. Isolated non-contact output (photo-MOS relay) No polarity. When comparator operating condition is ON, resistance between COMP and COM 50 Ω or less for C2 option (2 comparator) / 100 Ω or less for C1 option ( 1 comparator) Peak AC/DC280V 100mA (resistive load) / ON resistance 50Ω or less Comparison is made with polarity (0 is greater than minus) [Over] ON when set value or more / [Under] ON when set value or less Depends on analog output update time setting. Response time 1ms (max) |