Thermocouple meter HT-2401 [Temperature Meter]

Compatible with various thermocouple[TC] types K, J, R, E, T, B, N, S

Select a TC type and measure the temperature.
The electromotive force corresponding to the temperature difference between the tip of the TC and the main body of the instrument
is measured as a voltage and converted to a temperature.


Measurement -270°C to +1820°C
Measurement resolution 0.1℃/1℃

Rate setting between two points Linear correction formula

Thermocouples

Customization is possible, built into box, multi-CH, etc.     >Customization page
Can be calibrated as an embedded product.
Calibration documents and traceability charts can be issued.
19-inch rack mounting Built into box Multi-CH with indicator HDMI output + display software With SD logger
■Selectable measurement type and number of CHs ■Convenient, easy, low cost ■Customize input power and output signals to suit your facility
Output variety 2ms update
CAN output [Option]
3-wire wiring greatly reduces wiring compared to BCD
Powerful error detection system for improved noise immunity and safety

16bit analog output [Option]
Resolution 50,000 or more/Full scale and zero scale settings, output zero adjustment

Alarm output [Option] 
1 to 4points [Depends on model]

USB Type-C [standard feature] 
Data logging, setting, power supply

Reasonable and easy-to-use 1CH type  > Low-cost version
HT-2401 Display only, Equipped with USB communication 32,500 yen and up* [tax not included]
HT-2451 DIN rail mounting, no display 28,100 yen and up* [tax not included]
*Reference price in Japan.

Smart configuration and logging
Setting management, report printing, and CSV logging with dedicated software LINKSHIP [free of charge]  > LINKSHIP page

CSV logging Compare and match set value lists PC save and copy of set values Prevention of misoperation Quickly set up multiple units
Time-specified logging, such as 600 seconds, is also available. Custom logger software is also available. Multiple lists of set values can be compared and collated. It prevents errors in note-taking and manual entry, and reduces management effort and costs. Settings and confirmations can be made on the PC screen, preventing misoperation of the main unit's buttons. Even multiple units can be easily configured by simply writing the configuration data.

USB Type-C connector standard: configuration, logging, power supply   Setup unit without LINKSHIP or PC required:
Program Loader PL-5A
  [sold separately]

  PL-5A Product Page

*Setting only / data logging and power supply not available


Rear USB 2411 type also available.
 

Easy wiring and installation On-site usability considerations

Simply connect the wires to the connector and plug it in by hand.
Detachable Plug Connectors
Fixed Connectors
[Customized support]non-detachable
Mounting hooks:
Fixing the main body

Spring lock type [standard]

M2 screw type [J04 option]


*Not included in 2431 and 2451 types.

Various ways to use

Carry and measure Built in panel DIN rail mounting
Carry PC, instrument, and cable, and measure with USB power supply. Storage case BXS-1 is also available. 24x48 size contributes to downsizing One-touch fixture and display-less type are available.

BXS-1 Overview


HT-24 Dimensions
DIN rail attachment
[sold separately]

BXF-DN2 page
HT-2451
Display-less type
HT-2431
With display type

Applications   See more Applications > 

Integration into burn-in test equipment

Temperature
Pressure
Electric Current
Constant-temperature
oven
Temperature
Voltage
Current
Resistance
Frequency
HT
PSV
PSA
RR
SPD

CAN
USB
Analog output
RS-485
→→→

PC
Oscilloscopes
Data Loggers
Signal Input
Characteristics
Measurement

Good/Fail judgment
Alarm output
→→→

PLC

Fluid temperature difference measurement for flow error compensation
Temperature
Thermocouple meter

CAN
USB
Analog output
RS-485
→→→

PLCs, loggers, PCs, etc.



Flow rate
Speeedmeter/Counter
[Switchable]
Flow rate
Speeedmeter/Counter
[Switchable]
Temperature
Thermocouple meter

Temperature (thermocouple) - CAN conversion etc.

Specifications 

  Name
Model
Thermocouple meter
HT-2401
Input  Number of input points
1 point
  Measuring range Thermocouple types
K
J
R
E
T
B
N
S
Measuring range A [0.1°C]
-170.0 to 1370.0 °C
-200.0 to 1100.0 °C
100.0 to 1760.0 °C
-140.0 to 1000.0 °C
-240.0 to 400.0 °C
480.0 to 1820.0 °C
580.0 to 940.0 °C
450.0 to 1760.0 °C
Measuring range B [1°C]
-250 to 1370 °C
-200 to 1100 °C
-50 to 1760 °C
-260 to 1000 °C
-270 to 400 °C
100 to 1820 °C
-220 to 1300 °C
50 to 1760 °C
Accuracy
(±0.5% FS)
(±0.5% FS)
(±0.5% FS)
(±0.5% FS)
(±0.5% FS)
(±0.5% FS)
(±0.5% FS)
(±0.5% FS)
 


Measurement resolution
Burnout
Calibration
*Accuracy in parentheses ( ) is specified by setting output update time of 100 ms.
*Measurements can be made with a resolution of 0.1°C within the measurement range.
Cold junction compensator accuracy: ±2°C
Measurement range A: 0.1°C /Measurement range B: 1°C
display "bAhn" in 7seg characters
Adjustable with zero reference voltage input and span reference voltage input for each thermocouple type
(Adjustment command via CAN or USB command)
Calculation Moving Average Number
Hold function

Hold input
Additional control input
Setting value memory
1 to 99
Set from current value, max. value, min. value, max. fluctuation range, zero shift (hold / external control input)
Zero shift function, set the current input level as the zero reference value for measurement
ON when shorted to 1-point GND, OFF when open (hold function is set)
ON when shorted to 2-point GND, OFF when open (hold function assigned and set for each input)
Nonvolatile memory (EEPROM)
Display Display Color
Indicator (Numerical Display)

Decimal point position
Display update time
Display moving average times
Indicator
[Standard] HT-2401 Red LED / [Option] HT-2401-G Green LED
5-digit, 7-segment LED Character height 9mm / Display range:-19999 to 99999
Over indication: OL / Zero indication: Leading zero suppress
Semi-fixed Decimal point position is set in program mode. [0] *****,  [1] ****.* to [4] *.****
0.3 sec (settable from 0.1 to 9.9 sec)
1 time (settable from 1 to 9 times)  
LED 1 point (selected from power/ control input)
USB
communication
Communication specification
Communication function
Output update time
USB2.0 (Serial port communication by USB virtual COM 230.4kbps)
Writing/reading of set values/continuous output of measured values
2 to 9999ms, settable in 2ms increments
General
Information
Operating temp./humidity
Power supply voltage
Power consumption
Isolation
Withstand voltage
Weight
-10°C to +50°C / 35 to 85%RH (non-condensing)
7 to 60V DC with reverse connection protection. Power supply from USB host: 5V
3W or less
Power supply / Thermocouple input / Other input/output
Between each terminal of the above isolations: 500V DC between each terminal for 1 minute
Approx. 60g
CAN Option Communication specifications
Communication function
Setting item
Output update time
CAN2.0B 10 k / 20 k / 50 k / 100 k / 250 k / 500 k / 1M bps, [Format] INTEL/MOTOROLA, [Frame] 11bit/29bit
Writing/reading of set values/continuous output of measured values
CAN transmit ID, CAN receive ID
2 to 9999 ms (settable in 2 ms increments)
RS-485
communication
option
Number of connected units
Communication format
Communication method
Communication standard
Communication speed
Data format
Communication code
Communication function

32 units (max.)
2-wire multi-drop serial communication
Asynchronous
RS-485
9,600 /19,200 /38,400bps Set in program mode
Start bit: 1 bit / Stop bit: 1 bit / Data length: 8 bits / Parity bit: None
ASCII
Writing/reading set values/measurement value output (response to request command)
Modbus RTU support is available upon request.
Analog
output option
Output signals
D/A conversion method
Output resolution
Load resistance
Output accuracy
Temperature fluctuation
Output scaling
Output update time
Response time after calculation
[E1] 0 to 10 V, [E5] 0 to 5 V, [R1] ±10 V, [R5] ±5 V, [I1] 4 to 20 mA    [Select a model]
DAC conversion method
16bit (50,000 or more)
[Voltage output] 4.7kΩ or more / [Current output] 300Ω or less
[Voltage output] ±0.1% of FS @ 23°C / [Current output] ±0.1% of FS @ 23°C
±200ppm/℃ or less
Arbitrary scaling is possible by full scale and zero scale settings
2 to 9999ms settable in 2ms increments
1 ms or less, 0 - 90% response
Alarm
[Comparator]
output option
Setting method
Number of outputs
Output form

Output load voltage
Output logic



Update time
Set in program mode
2 points
Isolated non-contact output (photo-MOS relay) No polarity.
When comparator operating condition is ON, resistance between COMP and COM: 50 Ω or less

Peak AC/DC280V 100mA (resistive load) / ON resistance 50Ω or less
Comparison is made with polarity (0 is greater than minus)
[Over] ON when set value or more / [Under] ON when set value or less/
[Within range] ON within the set value, OFF outside the range
[Out of range] ON outside the set value, OFF inside the range

Depends on analog output update time setting. Response time 1ms (max)

To the page of Thermometers list

To the page top